Arterial Line - Square Wave Test 
The arterial line ...
17.6K
Description

Arterial Line - Square Wave Test 

The arterial line can measure BP inaccurately unless properly calibrated. Rapidly flushing the line (by pulling the release on the flush device) generates a square wave. Counting oscillations after the square wave indicates if the arterial line is working properly.

Normal = accurate BP

    - 1.5 - 2 oscillations

OVERdamped = falsely LOW BP

    - <1.5 oscillations

    - Absence of dicrotic notch

    - Fix: Remove extra tubing & any air bubbles.

UNDERdamped = falsely HIGH BP

    - >2 oscillations

    - Additional artifactual spikes

    - Fix: Adjust filter settings on monitor



Nick Mark MD @nickmmark



#squarewave #test #dampening #damped #arterial #artline #diagnosis 
Contributed by

Dr. Gerald Diaz
@GeraldMD
Board Certified Internal Medicine Hospitalist, GrepMed Editor in Chief πŸ‡΅πŸ‡­ πŸ‡ΊπŸ‡Έ - Sign up for an account to like, bookmark and upload images to contribute to our community platform. Follow us on IG:  https://www.instagram.com/grepmed/ | Twitter: https://twitter.com/grepmeded/
Medical jobs
view all

0 Comments

Related content